    Use of second-order calibration for residue screening of some triazines in the presence of coeluting interferences by gas chromatography-selected ion mass spectrometry

    , Article Analytica Chimica Acta ; Volume 537, Issue 1-2 , 2005 , Pages 89-100 ; 00032670 (ISSN) Jalali Heravi, M ; Vosough, M ; Sharif University of Technology
    Elsevier  2005
    Abstract
    The quantities of residues of some triazines such as prometon, propazine, atrazine and simazine in complex matrices of apple samples were determined, using gas chromatography-selected ion mass (GC-SIM) spectrometry. Generalized rank annihilation method (GRAM) as a second-order calibration technique was used for screening, resolving and finally determining the amounts of the residues. Before the GRAM analysis, different steps of data preprocessing such as background correction, de-skewing and standardization for rank alignment was used for every target analyte. The de-skewing and rank alignment algorithms were used for bilinearity and trilinearity corrections, respectively. The two data...