    Potentiometric membrane sensors based on zirconyl(IV) phthalocyanine for detection of sulfosalicylic acid

    , Article Talanta ; Volume 63, Issue 2 , 2004 , Pages 371-376 ; 00399140 (ISSN) Shahrokhian, S ; Taghani, A ; Hamzehloei, A ; Mousavi, S. R ; Sharif University of Technology
    Elsevier  2004
    Abstract
    A metallophthalocyanine complex with zirconium(IV) ion in the center (as an oxo-zirconium, Zr=O, group) was used in poly vinyl chloride (PVC) membranes for the selective detection of 5-sulfosalicylic acid (SSA). The resulting electrodes demonstrate Nernstian responses over a wide range of sulfosalicylic acid concentration (10-6 to 10-1 mol dm-3) with a slope of about -29 mV per decade. The influence of lipophilic ion-exchanger sites on the response properties of the electrodes was investigated. The optimal potentiometric response was observed for the electrode in the presence of about 150 mol% of cationic additive (relative to ionophore) in the phase membrane. The electrodes have a fast...